Calibratori di pressione digitali
ADT672
Applicazioni
• Pressure ranges to 36,000 psi (2500 bar)
• Hart Communication capability
• Measure mA or V, and with 24V loop power
• Easy-to-use, inexpensive pressure calibrator with uncertainty better than 0.025%FS
• Gauge calibration
• Precision pressure measurement
• Pressure transmitters calibration
• Pressure switch testing
• Safety relief valve resting
• Pressure regulator testing
• Hydraulic leak testing
SpecificheADT 672-02: 0.025%FS
ADT 672-05: 0.05%FS
• Pressure ranges to 36,000 psi (2500 bar)
• Measure mA or V with 0.02% RD + 0.003%FS accuracy
• Power transmitters during test using 24V loop supply
• Pressure switch test
• Hart Communication capability
• Advanced temperature compensation
• Dual readout
• Min/Max/Hold to capture changing measurements
• Data logging
• Large, easy to read display with 6-digit resolution
• Back lighted display
• Rechargeable battery or AC adapter
• NIST traceable calibration (optional)
• Gauge calibration
• Precision pressure measurement
• Pressure transmitters calibration
• Pressure switch testing
• Safety relief valve resting
• Pressure regulator testing
• Hydraulic leak testing
Gauge Pressure [1]
|
|||||
P/N
|
Pressure Range
(psi)
|
Pressure Range
(bar)
|
Media [2]
|
Accuracy
(% FS)
|
Burst Pressure
|
V15
|
-15
|
-1.0
|
G
|
0.025 (0.05)
|
3 X
|
GP2
|
2
|
0.16
|
G
|
0.05
|
3 X
|
GP5
|
5
|
0.35
|
G
|
0.025 (0.05)
|
3 X
|
GP10
|
10
|
0.7
|
G
|
0.025 (0.05)
|
3 X
|
GP15
|
15
|
1.0
|
G , L [3]
|
0.025 (0.05)
|
3 X
|
GP30
|
30
|
2.0
|
G , L [3]
|
0.025 (0.05)
|
3 X
|
GP50
|
50
|
3.5
|
G , L
|
0.025 (0.05)
|
3 X
|
GP100
|
100
|
7.0
|
G , L
|
0.025 (0.05)
|
3 X
|
GP300
|
300
|
20
|
G , L
|
0.025 (0.05)
|
3 X
|
GP500
|
500
|
35
|
G , L
|
0.025 (0.05)
|
3 X
|
GP1K
|
1000
|
70
|
G , L
|
0.025 (0.05)
|
3 X
|
GP3K
|
3000
|
200
|
G , L
|
0.025 (0.05)
|
3 X
|
GP5K
|
5000
|
350
|
G , L
|
0.025 (0.05)
|
3 X
|
GP10K
|
10000
|
700
|
G , L
|
0.025 (0.05)
|
2 X
|
GP15K
|
15000
|
1000
|
G , L
|
0.1
|
2 X
|
GP25K
|
25000
|
1600
|
G , L
|
0.1
|
1.5 X
|
GP30K
|
30000
|
2000
|
G , L
|
0.1
|
1.5 X
|
GP36K
|
36000
|
2500
|
G , L
|
0.1
|
1.5 X
|
Absolute Pressure
|
|||||
P/N
|
Pressure Range
(psi)
|
Pressure Range
(bar)
|
Media
|
Accuracy
(% FS)
|
Burst Pressure
|
AP5
|
5
|
0.35
|
G
|
0.1 (0.2)
|
3 X
|
AP10
|
10
|
0.7
|
G
|
0.1 (0.2)
|
3 X
|
AP15
|
15
|
1.0
|
G
|
0.1 (0.2)
|
3 X
|
AP30
|
30
|
2.0
|
G
|
0.1 (0.2)
|
3 X
|
AP50
|
50
|
3.5
|
G
|
0.1 (0.2)
|
3 X
|
AP100
|
100
|
7.0
|
G , L
|
0.1 (0.2)
|
3 X
|
AP300
|
300
|
20
|
G , L
|
0.1 (0.2)
|
3 X
|
AP500
|
500
|
35
|
G , L
|
0.1 (0.2)
|
3 X
|
AP1K
|
1000
|
70
|
G , L
|
0.1 (0.2)
|
3 X
|
AP3K
|
3000
|
200
|
G , L
|
0.1 (0.2)
|
3 X
|
AP5K
|
5000
|
350
|
G , L
|
0.1 (0.2)
|
3 X
|
Differential Pressure
|
|||||
P/N
|
Pressure Range
(inH2O)
|
Pressure Range
(mbar)
|
Media
|
Accuracy
(% FS)
|
Burst Pressure
|
DP1
|
1
|
2.5
|
G
|
0.4
|
100 X
|
DP2
|
2
|
6.0
|
G
|
0.2
|
100 X
|
DP5
|
5
|
10
|
G
|
0.1
|
50 X
|
DP10
|
10
|
25
|
G
|
0.1
|
20 X
|
DP20
|
20
|
50
|
G
|
0.05
|
20 X
|
DP30
|
30
|
100
|
G
|
0.05
|
20 X
|
Compound Pressure
|
|||||
P/N
|
Pressure Range
(psi)
|
Pressure Range
(bar)
|
Media
|
Accuracy
(%FS)
|
Burst Pressure
|
CP2
|
±2
|
±0.16
|
G
|
0.05 |
3X
|
CP5
|
±5
|
±0.35
|
G
|
0.025(0.05)
|
3X
|
CP10
|
±10
|
±0.7
|
G
|
0.025(0.05)
|
3X
|
CP15
|
±15
|
±1
|
G
|
0.025(0.05)
|
3X
|
CP30
|
-15 to 30
|
-1 to 2
|
G
|
0.025(0.05)
|
3X
|
Electrical Measurement and Source Accuracy
|
||
|
Range
|
Accuracy
|
Voltage DC
|
±30.0000V
|
± (0.02%RD+0.003%FS)
|
Current DC
|
±30.0000mA
|
± (0.02%RD+0.003%FS)
|
DC 24V
|
24V±0.5V max:50mA Protect at: 120mA
|
|
Switch [1]
|
Status OPEN/CLOSED
|
Mode: manual and automatic
Hourly-record: record the data every hour
Interval-record: set by user
External power: 110V/220V power adapter(10V DC)
Data length: 8 bits
Stop bit: 2 bits
Address: from 1 to 112
2 pieces 1.5-meter test leads
2 pieces 0.156 inch (Ø4 mm) test hose (for differential pressure gauge only)
Model number
|
Description
|
9702
|
Spare rechargeable Li-ion polymer battery for ADT 672
|
9814
|
Spare 110V/220V external power adapter (DC 10V) for 672 digital pressure calibrator
|
9902
|
Carrying case for digital pressure gauge (up to 4 pieces ADT 672)
|
9021
|
Spare 2 pieces test leads (1.5-meter) for ADT 672
|
9050
|
USB to RS232 (DB9/M) Adapter
|
9111
|
NIST traceable calibration for ADT 672 series gauges (≤10,000psi)
|
9112
|
NIST traceable calibration for ADT 672 series gauges (>10,000psi)
|
9500
|
Additel/Log software for downloading ADT672 test results stored in memory
|
9502
|
Additel/Log II real time logging and graphical software for ADT681 and ADT672
|